1. Quality of test design/production. There is significant variation of performance between tests, users are advised to check test performance (see WHO testing information available from [https://www.who.int/teams/global-malaria-programme/case-management/diagnosis/rapid-diagnostic-tests].</br> | 1. Quality of test design/production. There is significant variation of performance between tests, users are advised to check test performance (see WHO testing information available from [https://www.who.int/teams/global-malaria-programme/case-management/diagnosis/rapid-diagnostic-tests].</br> | ||
2. Lot to lot variability. Variability between batches has been observed and should be considered if RDT performance is less than expected.</br> | 2. Lot to lot variability. Variability between batches has been observed and should be considered if RDT performance is less than expected.</br> | ||
'''2. Heat stability of the RDT test''' storage</b> | |||
RDT tests can suffer degradation if stored in inappropriate heat or humidity. Generally tests are stable in the range 2–30°C or greater. However these limits may be exceeded in many countries and manufacturers recommend continuous product refrigeration to the point of use. This may create problems for developing countries where cold-chain continuity cannot be guaranteed. | |||
'''3. Age of test or reagents'''</br> | |||
RDTs typically have a shelf-life of 18–24 months, which inludes time for distribution and use, providing storage is appropriate. | |||
Operator-Specific Factors | Operator-Specific Factors | ||
Operator-Inappropriate placement of reagent or blood drop | Operator-Inappropriate placement of reagent or blood drop | ||
| Line 26: | Line 29: | ||
In order to improve test sensitivities when dealing with non-ideal biomarkers, manufacturers of RDTs often seek to use the multiplexed detection of distinct, species-specific antigens. In the case of malaria, RDTs are often multiplexed to detect HRP2 and the P. vivax variant of parasite lactate dehydrogenase (Pv-pLDH), to distinguish between infections by P. falciparum and P. vivax. Alternatively, tests can be multiplexed to select for Pf-pLDH and HRP2, which yields highly sensitive indications for falciparum malaria. [16] Multiplexed RDTs have also proven to be efficacious in discriminating between distinct diseases with non-specific symptomatic presentations | In order to improve test sensitivities when dealing with non-ideal biomarkers, manufacturers of RDTs often seek to use the multiplexed detection of distinct, species-specific antigens. In the case of malaria, RDTs are often multiplexed to detect HRP2 and the P. vivax variant of parasite lactate dehydrogenase (Pv-pLDH), to distinguish between infections by P. falciparum and P. vivax. Alternatively, tests can be multiplexed to select for Pf-pLDH and HRP2, which yields highly sensitive indications for falciparum malaria. [16] Multiplexed RDTs have also proven to be efficacious in discriminating between distinct diseases with non-specific symptomatic presentations. | ||
Revision as of 17:35, 12 September 2024
Navigation
Go Back
| The value of RDT compared with microscopy: thick blood films
see also: [1] Selection and procurement of MDTs 2. Heat stability of the RDT test storage RDT tests can suffer degradation if stored in inappropriate heat or humidity. Generally tests are stable in the range 2–30°C or greater. However these limits may be exceeded in many countries and manufacturers recommend continuous product refrigeration to the point of use. This may create problems for developing countries where cold-chain continuity cannot be guaranteed. 3. Age of test or reagents
In the case of malarial RDTs, the WHO requires a panel detection score of 75% at a parasitaemia of 200 parasites/μl in order for the product to warrant endorsement. [16] However, the threshold for pyrogenic onset has been shown to vary from 10 to 200,000 parasites/μl, with 22% of patients developing their first acute fever at a parasite load below 200/μl. [75] Thus, even for the select assortment of commercial malaria RDTs demonstrating compliance with WHO standards, sufficiently low limits of detection have not been demonstrated for the early and accurate diagnosis of nascent infections, or of asymptomatic carriers.
|